Agilent Mark-VI Spray Chamber Application Note for Flame Atomic Absorption Spectrometry

The document introduces the performance of Agilent Mark-VI spray chamber, including characteristic concentration, drainage, carbon deposition, burner blockage and interference effects


Agilent Mark-VI Spray Chamber Application Note for Flame Atomic Absorption Spectrometry - cover page
Brand
Evaluation
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
288 KB
Published
12 June, 2012
Updated
30 September, 2023
MD5 checksum
1E460D333686282D36555E167C70F4F8

About this manual

The document describes the design features of the Agilent Mark-VI spray chamber, including its pneumatic nebulizer, adjustable glass bead, twin-head mixing paddle, steeply angled floor, and wide-bore drainage tube. These components are designed to minimize carbon build-up, prevent burner blockage, and ensure rapid sample clear-out.

What's inside

  • Introduction
  • Description of Design
  • Drainage
  • Carbon Build-Up
  • Burner Blockage From High-Solids Solutions

Frequently asked questions

How quickly does the Mark-VI spray chamber drain a sample solution?

The drainage is complete within 5 seconds after aspirating the blank.

Download PDF (288 KB)