This document describes the features and application of the Agilent Mercury Concentration Accessory (MCA-90). The accessory allows for the determination of mercury at ultra-low levels and is suitable for environmental monitoring. By utilizing the amalgamation of mercury with gold, the MCA-90 enhances the sensitivity of mercury signals for measurement.


Agilent MCA-90 Mercury Concentration Accessory Application Note - cover page
Brand
Evaluation
Category
Other Electronics
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
100 KB
Published
12 June, 2012
Updated
29 September, 2023
MD5 checksum
70ECE7A6294FDC5C2B0F801E0F8E6E47

About this manual

The Agilent MCA-90 Mercury Concentration Accessory is a microprocessor-based device designed to enhance the detection of mercury at ultra-low levels when used with the Agilent VGA-76 Vapor Generation Accessory and SpectrAA series atomic absorption spectrometers. It utilizes a gold collecting surface to trap mercury vapor, which is then electrically heated to release adsorbed mercury for measurement.

The document details a study evaluating the MCA-90 using the US EPA digestion procedure for environmental samples. Results show that with a 40-second collection time, the system achieves a detection limit of 0.01 pg/L (10 ppt), significantly lower than the VGA-76 alone. The note also provides practical guidelines for setting collection times, delay times, and heat times to optimize sensitivity and sample throughput.

What's inside

  • Introduction
  • Experimental Methodology
  • Results
  • Stability Study
  • Practical Operation
  • Table 1. EPA Digestion Scheme
  • Table 2. Instrument Parameters
  • Table 6. Collection Time Versus Analytical Sensitivity

Specifications

Detection Limit0.01 pg/L (10 ppt)
Collection Time (Manual Mode)40 seconds
Delay Time (Manual Mode)35 seconds
Heat Time45 seconds
Wavelength253.7 nm
Sample Volume (40s collect)16 mL

Frequently asked questions

What is the detection limit of the MCA-90 with a 40-second collection time?

The detection limit is 0.01 pg/L (10 ppt).

How does the MCA-90 enhance mercury signals?

It uses a gold collecting surface to trap mercury vapor, which is then electrically heated to release the adsorbed mercury for measurement.

Which instrument series is the MCA-90 compatible with?

It is a microprocessor-based accessory for the SpectrAA series.