This manual describes the maintenance of the Agilent 6100 series single quadrupole LC/MS system
This document serves as a maintenance guide for the Agilent 6100 Series Single Quad LC/MS System. It details specific tasks required to keep the instrument in optimal working condition, including daily and weekly cleaning procedures for various ion source components.
The manual covers step-by-step instructions for flushing the nebulizer, cleaning electrospray and APCI spray chambers, replacing needles, adjusting needle positions, and maintaining the desolvation assembly. It also includes procedures for checking calibration levels, replacing LC filter elements, and performing bakeouts on the system. This manual is written in Japanese.
The nebulizer should be flushed daily or at the end of each shift when using small amounts of sample and buffer.
The chamber should be cleaned daily or at the end of each shift, or whenever contamination accumulation is suspected after changing samples or analyses.