Agilent MMI Tutorial Manual

This document introduces the application and advantages of Agilent Technologies' Agilent Multimode Inlet (MMI) for large volume injection (LVI). The LVI method allows for increased sample injection volume, improved analysis results, and reduced solvent usage. Additionally, the LVI process can be automated for higher reproducibility.


Agilent MMI Tutorial Manual - cover page
Brand
Agilent
Document type
User Manual
Language
Japanese
Pages
19
File format
PDF
File size
3.2 MB
Published
08 June, 2012
Updated
28 September, 2023
MD5 checksum
099DEEA9009827E19FD03DEE8622BA56

About this manual

The document serves as a tutorial for the Agilent Multimode Inlet (MMI). It introduces Large Volume Injection (LVI) methods to improve analytical sensitivity by injecting larger sample volumes compared to traditional methods. The text explains that LVI allows researchers to increase sample mass reaching the detector, thereby improving signal-to-noise ratios and lowering detection limits.

The guide details three specific operating modes supported by the MMI: Hot Splitless, Cold Splitless, and Solvent Vent. It describes the operational principles of each mode, including temperature programming requirements and liner interactions. The document concludes with a step-by-step tutorial demonstration using an FID MDL check-out sample to illustrate method development. This manual is written in Japanese.

What's inside

  • Large Volume Injection (LVI)
  • Hot Splitless
  • Cold Splitless
  • Solvent Vent Mode
  • Sample Introduction Methods
  • Detector Sensitivity
  • Method Development Demo
  • FID MDL Check-out Sample

Frequently asked questions

What is the maximum injection volume supported by this method?

The Large Volume Injection (LVI) method can be applied to injection volumes ranging from a few microliters up to more than 1 milliliter.

What is the recommended column type and dimensions for this tutorial?

The tutorial specifies using an HP-5 column with dimensions of 30 m x 0.32 mm x 0.25 um.

What is the required inlet temperature for the splitless mode in Step 1?

The inlet temperature must be set to 250 ?C during the initial splitless analysis step.