This document details the replacement method of the inlet/waste tubing of the fraction collector
This document provides step-by-step instructions for replacing contaminated or damaged inlet/waste tubings and valve to needle tubings on the Agilent G1390A Universal Interface Box (UIB). The procedures cover safety precautions, required tools, and specific installation steps for both preparative scale (0.8 mm ID) and analytical scale (0.25 mm ID) tubing kits.
| Inlet / waste tubing assembly ID (Preparative scale) | 0.8 mm ID |
|---|---|
| Inlet / waste tubing assembly ID (Analytical scale) | 0.25 mm ID |
Regularly inspect the tubings and exchange them if they are worn out or show visible signs of damage.
The tubing should typically be exchanged once every year.
Use PN G1364-68711 for the preparative scale kit or PN G1364-68712 for the analytical scale kit.