Describes the conditions required for installation of Varian 210-MS, 220-MS, and 225-MS GC/MS Ion Trap Mass Spectrometer, including laboratory safety requirements, personnel preparation, workbench space, temperature, humidity, power and gas
This document provides pre-installation instructions for the Agilent 210-MS, 220-MS, and 225-MS GC/MS Ion Trap Mass Spectrometers. It outlines a checklist of requirements that must be met before requesting installation, including site safety regulations, entrance clearance, bench space, and environmental conditions.
The optimal operating temperature is between 16 and 30 °C (61-86 °F).
Relative humidity must be maintained between 20% and 80%, with no condensation.
The entrance to the laboratory must be at least 92 cm (36 in.) wide.