Omron HEM-7118 Error E

Omron HEM-7118 Electronic Blood Pressure Monitor · taken from the manufacturer’s manual, pages cited

Error codeE

The code E indicates body movement during measurement, air hose connector detached, or cuff wrapping method is incorrect. The manual's first action is: “Retake the measurement. Keep the body still and do not speak during the measurement.”

Model
Omron HEM-7118
Display shows
E
Type
Error code
Source
Manual p. 2

What E means

Body movement during measurement. p. 2

Common causes

  • Body movement during measurement. p. 2
  • Air hose connector detached. p. 2
  • Cuff wrapping method is incorrect. p. 2
  • Clothing interferes with the cuff. p. 2
  • Cuff leakage. p. 2
  • Cuff pressure has exceeded 299 mmHg (39.9 kPa). p. 2

What to do

  1. Retake the measurement. Keep the body still and do not speak during the measurement. p. 2
  2. Insert the air connector correctly. Refer to “3.1 Applying the Cuff”. p. 2
  3. Wrap the cuff correctly. Refer to “3.1 Applying the Cuff”. p. 2
  4. Remove clothing that interferes with the cuff. p. 2
  5. Replace with a new cuff. p. 2
  6. Do not inflate the cuff above 299 mmHg (39.9 kPa). Refer to “3.3 Explanation of Special Situations”. p. 2

Before you service the product, read the safety instructions in the manual p. 1.

If E is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Omron service.

This page summarises what the manufacturer’s manual says about E; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Omron. Found a mistake? Report an error.