Modulation calrestorefailure means there is either a failure in writing data to RAM or in reading data from the calibration ROM. The manual's first action is: “Replace the A3 assembly. Both the calibration ROM and RAM are on the A3 assembly.”
- Model
- Hewlett-Packard 8648A/B/C/D
- Display shows
- Modulation calrestorefailure (504)
- Type
- Error code
- Source
- Manual p. 223
What 504 means
Modulation calrestorefailure p. 223
Common causes
- There is either a failure in writing data to RAM or in reading data from the calibration ROM. p. 223
What to do
- Replace the A3 assembly. Both the calibration ROM and RAM are on the A3 assembly. p. 223
Before you service the product, read the safety instructions in the manual p. 277.
If 504 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Hewlett-Packard service.
This page summarises what the manufacturer’s manual says about 504; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Hewlett-Packard. Found a mistake? Report an error.