Bruker IPSO 19" Error C0: Try to boot

NMR spectrometer · taken from the manufacturer’s manual, pages cited

Error codeC0

The system boots successfully but reports 'No Operating System found' due to a possible network or boot configuration issue. The manual's first action is: “1.Check Ethernet connection. Yellow LINK LED on? Green Rx/Tx LED active? 2.Check in BIOS if Netboot=yes? (see below) or Netboot is at the top of the list in the submenu 'Boot Device Priority' and set the item "Onboard LAN PXL ROM" to "Enabled"”

Model
Bruker IPSO 19"
Display shows
Try to boot (C0)
Type
Error code
Source
Manual p. 30

What C0 means

Try to boot p. 30

Common causes

  • Successful BIOS process but "No Operating System found" p. 30

What to do

  1. 1.Check Ethernet connection. Yellow LINK LED on? Green Rx/Tx LED active? 2.Check in BIOS if Netboot=yes? (see below) or Netboot is at the top of the list in the submenu 'Boot Device Priority' and set the item "Onboard LAN PXL ROM" to "Enabled" p. 30

If C0 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bruker service.

This page summarises what the manufacturer’s manual says about C0; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bruker. Found a mistake? Report an error.