Bruker IPSO 19" Error 49: Unsuccessful PCI configuration

NMR spectrometer · taken from the manufacturer’s manual, pages cited

Error code49

PCI configuration failed due to a faulty PC-Module or defective Tx- or RxController in the system. The manual's first action is: “1.Check insertion of the PC-Module 2.Remove the controllers one after the other and try again”

Model
Bruker IPSO 19"
Display shows
Unsuccessful PCI configuration (49)
Type
Error code
Source
Manual p. 30

What 49 means

Unsuccessful PCI configuration p. 30

Common causes

  • Faultily inserted PC-Module (Host Controller) p. 30
  • Any defective Tx- or RxController in the system p. 30

What to do

  1. 1.Check insertion of the PC-Module 2.Remove the controllers one after the other and try again p. 30
  2. 3.Check voltages of the Power Supply 4.Exchange the PC-Module p. 30

If 49 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bruker service.