The system stops at B0 with 2 beeps because the timer contains corrupted time and date information. The manual's first action is: “To recover the content of the timer: 1. Press the RESET button or 2. Connect an USB-keyboard and press F1 to resume and correct time and date with LINUX or 3. Connect a monitor too, press F2 to enter the BIOS setup and correct the time and date or set 'Hold on errors' to "NO"”
- Model
- Bruker IPSO 19"
- Display shows
- Check for errors, stops at B0 with 2 beeps in case of error (B0)
- Type
- Error code
- Source
- Manual p. 30
What B0 means
Check for errors, stops at B0 with 2 beeps in case of error p. 30
Common causes
- The timer contains corrupted time and date information. p. 30
What to do
- To recover the content of the timer: 1. Press the RESET button or 2. Connect an USB-keyboard and press F1 to resume and correct time and date with LINUX or 3. Connect a monitor too, press F2 to enter the BIOS setup and correct the time and date or set 'Hold on errors' to "NO" p. 30
If B0 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bruker service.
This page summarises what the manufacturer’s manual says about B0; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bruker. Found a mistake? Report an error.
