Bruker IPSO 19" Error B0: Check for errors, stops at B0 with 2 beeps in case of error

NMR spectrometer · taken from the manufacturer’s manual, pages cited

Error codeB0

The system stops at B0 with 2 beeps because the timer contains corrupted time and date information. The manual's first action is: “To recover the content of the timer: 1. Press the RESET button or 2. Connect an USB-keyboard and press F1 to resume and correct time and date with LINUX or 3. Connect a monitor too, press F2 to enter the BIOS setup and correct the time and date or set 'Hold on errors' to "NO"”

Model
Bruker IPSO 19"
Display shows
Check for errors, stops at B0 with 2 beeps in case of error (B0)
Type
Error code
Source
Manual p. 30

What B0 means

Check for errors, stops at B0 with 2 beeps in case of error p. 30

Common causes

  • The timer contains corrupted time and date information. p. 30

What to do

  1. To recover the content of the timer: 1. Press the RESET button or 2. Connect an USB-keyboard and press F1 to resume and correct time and date with LINUX or 3. Connect a monitor too, press F2 to enter the BIOS setup and correct the time and date or set 'Hold on errors' to "NO" p. 30

If B0 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bruker service.