On the Bio-Rad PowerPac HV, code 18 indicates a temperature error with the run paused when the temperature remains 5°C less than the set temperature for more than 15 minutes, which can be caused by a detached temperature probe, an insufficient power setting, or a temperature probe no longer connected to or making poor contact with the electrophoresis cell. The manual's first action is: “The run is paused and error code 18 is displayed in case of failure.”
- Model
- Bio-Rad PowerPac HV
- Display shows
- TEMPERATURE ERROR. CORRECT AND CONTINUE RUN PAUSED (18)
- Type
- Fault code
- Source
- Manual p. 17
What 18 means
TEMPERATURE ERROR. CORRECT AND CONTINUE RUN PAUSED p. 17
Common causes
- A failure occurs when the temperature remains 5°C less than the set temperature for more than 15 minutes. p. 17
- This can be caused by a detached temperature probe or an insufficient power setting. p. 17
- The set temperature cannot be reached because: 1. The temperature probe is no longer connected to the electrophoresis cell or the temperature probe made poor contact with the electrophoresis cell. p. 17
- 2. The programmed power is insufficient to reach the set temperature. p. 17
What to do
- The run is paused and error code 18 is displayed in case of failure. p. 17
- 1. Make sure the temperature probe is securely fastened to the electrophoresis cell. p. 17
- 2. Adjust the power so it is sufficient to reach the set temperature. p. 17
- Press CONTINUE after the appropriate corrections are made. p. 17
- Press CANCEL to terminate the run. p. 17
If 18 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.
This page summarises what the manufacturer’s manual says about 18; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.