On the Bio-Rad PowerPac HV, code 07 indicates run power failure from an AC power interruption during a run, with the run terminated because power failure detection is OFF. The manual's first action is: “• To continue a run after a power failure, activate the power failure detection mode in the Setup window prior to each run. See Section 3 for detailed explanation.”
- Model
- Bio-Rad PowerPac HV
- Display shows
- RUN POWER FAILURE (07)
- Type
- Error code
- Source
- Manual p. 38
What 07 means
RUN POWER FAILURE p. 38
Common causes
- AC power interruption during a run and the run is terminated because the Power failure detection is OFF p. 38
What to do
- • To continue a run after a power failure, activate the power failure detection mode in the Setup window prior to each run. See Section 3 for detailed explanation. p. 38
If 07 is still not resolved
The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.
This page summarises what the manufacturer’s manual says about 07; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.