Bio-Rad PowerPac HV Error 01: NOLOAD DETECTED

Electrophoresis power supply · taken from the manufacturer’s manual, pages cited

Error code01

On the Bio-Rad PowerPac HV, code 01 indicates no load detected because the electrophoresis cell is not connected to the power supply or power levels are too low, and the current load is less than 0.02 mA. The manual's first action is: “1. Make sure all electrical connections are making good contact and the cables and wire electrodes are in good shape.”

Model
Bio-Rad PowerPac HV
Display shows
NOLOAD DETECTED (01)
Type
Error code
Source
Manual p. 38

What 01 means

NOLOAD DETECTED p. 38

Common causes

  • 1. Electrophoresis cell not connected to the power supply or power levels too low. p. 38
  • 2. The current load is less than 0.02 mA. p. 38

What to do

  1. 1. Make sure all electrical connections are making good contact and the cables and wire electrodes are in good shape. p. 38
  2. 2a. Verify buffer levels are appropriate. p. 38
  3. 2b. Verify the electrophoresis application power requirements match PowerPac HV output range. See Section 3 to disable this safety feature. p. 38

If 01 is still not resolved

The manual gives no further instruction for this code. Open the code table in the PDF or contact Bio-Rad service.

This page summarises what the manufacturer’s manual says about 01; each statement links to the page it comes from, and quoted lines are copied word for word. It is not a substitute for the manual or for a qualified installer. ManualLib is not affiliated with Bio-Rad. Found a mistake? Report an error.