Perkin Elmer LS-50B Manuals

ManualLib has 2 documents for the Perkin Elmer LS-50B: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
LS-50B
Industry
Health
Documents
2
Languages
EN
Last added
2026-11-06

Start here

Perkin Elmer LS-50B Luminescence Spectrometer User’s Guide

Perkin Elmer LS-50B Luminescence Spectrometer User’s Guide

User Manual 208 pages · 5.2 MB · EN

This manual provides detailed instructions for safely operating and maintaining the LS-50B Luminescence Spectrometer. It covers installation procedures, accessory setup, safety precautions, and technical commands for system control. The guide also includes specifications, maintenance tips, and troubleshooting information for optimal instrument performance.

The LS-50B is a computer-controlled ratioing luminescence spectrometer capable of measuring fluorescence, phosphorescence, chemiluminescence, and bioluminescence. It features a xenon discharge lamp, gated photomultipliers, and Monk-Gillieson monochromators for precise wavelength control and spectral analysis.

Read online

Perkin Elmer LS-50B specifications

PrincipleComputer controlled ratioing luminescence spectrometer
SourceXenon discharge lamp, equivalent to 20 kW for 8 µs duration
Sample detectorGated photomultiplier with modified S5 response for operation up to around 650 nm
Wavelength accuracy+ 1.0 nm
Wavelength reproducibility+ 0.5 nm
Excitation range200-800 nm with zero order selectable
Emission range200-650 nm with standard photomultiplier, 200-900 nm with optional R928 photomultiplier

Source: Perkin Elmer LS-50B Luminescence Spectrometer User’s Guide

All Perkin Elmer LS-50B documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What are the wavelength accuracy and reproducibility specifications?

The wavelength accuracy is ±1.0 nm and the reproducibility is ±0.5 nm.

From: Perkin Elmer LS-50B Luminescence Spectrometer User’s Guide

What environmental conditions are required for the safe operation of the LS-50B?

The instrument operates correctly indoors at ambient temperatures between 15 °C and 35 °C with a relative humidity of 75% maximum, non-condensing.

From: Perkin Elmer LS-50B Luminescence Spectrometer User’s Guide

What type of cells are required for the LS50B spectrometer?

The instrument requires 1.0 cm path length cells that must be transparent on all four sides.

From: PerkinElmer LS50B Luminescence Spectrometer SOP

What are the default parameters for an emission scan?

The default settings include a start of 340, end of 480, emission of 388, Ex Slit of 5, Em Slit of 2.5, and Scan Speed of 50.

From: PerkinElmer LS50B Luminescence Spectrometer SOP