Micsig DP750 Manuals

ManualLib has 2 documents for the Micsig DP750 (Differential Probes): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
Micsig
Model
DP750
Industry
Electronics
Documents
2
Languages
EN · ZH
Last added
2026-11-16

Start here

Micsig DP750 High Voltage Differential Probe User Manual

Micsig DP750 High Voltage Differential Probe User Manual

User Manual 2 pages · 601 KB · EN

The Micsig DP750 is a high voltage differential probe with a built-in shielded test line and compact interface box. It provides a maximum 100 MHz bandwidth, a maximum input voltage of 750 V, switchable 50X/500X attenuation, one-press zero calibration, and an overvoltage alarm.

The manual covers the probe's specifications, panel description, operation steps, precautions, warranty, and safety. Applications listed in the text include research and development, commissioning, or maintenance of switching power supplies, frequency converters, electronic ballasts, frequency conversion home appliances, and other electrical power devices.

Read online

Micsig DP750 specifications

Bandwidth100 MHz (DP750-100); 50 MHz (DP750-50)
Rise Time≤ 3.5 ns (DP750-100); ≤ 7 ns (DP750-50)
Attenuation Ratio50X; 500X
Gain Accuracy±2%
Max Differential Test Voltage (DC+AC PK)50X: 75 V; 500X: 750 V
Max Common Mode Input VoltageCAT II 600 V
Input Referred Noise≤ 12 mVrms (50X); ≤ 120 mVrms (500X)
Common Mode Rejection Ratio (CMRR)> 80 dB (DC); > 80 dB (100 kHz); > 60 dB (1 MHz)
Differential Input Impedance8 MΩ || 1.25 pF
Output Voltage≤ 1.5 V
Power Rate1 W
Power SupplyMicsig UPI Interface; USB (Adaptor)
DimensionsPower Supply Box: 80*37*22 mm; Signal In/Output Box: 49*25*12.5 mm
Working Humidity10%-85%

Source: Micsig DP750 High Voltage Differential Probe User Manual

All Micsig DP750 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

How do I perform zero calibration on the DP750?

Short-circuit the two input terminals, power on, press and hold the Zero button for 3 seconds; the two indicators flash and calibration succeeds after 3 seconds.

From: Micsig DP750 High Voltage Differential Probe User Manual

What oscilloscope bandwidth is required?

The oscilloscope bandwidth must be no less than 100 MHz.

From: Micsig DP750 High Voltage Differential Probe User Manual

How do I calibrate the zero point of the probe?

Short the two input terminals, power on, and long press the 'Zero' button for 3 seconds until the two indicator lights flash simultaneously.

From: Micsig DP750 Differential Probe User Manual

What are the oscilloscope requirements for connecting the differential probe?

The oscilloscope bandwidth must be no lower than 100MHz, and the channel input impedance must be no lower than 1MΩ.

From: Micsig DP750 Differential Probe User Manual