Licel R9880U Manuals

ManualLib has 2 documents for the Licel R9880U: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
Licel
Model
R9880U
Documents
2
Languages
EN
Last added
2023-10-16

Start here

Licel R9880U Gated Photomultiplier Module User Manual

Licel R9880U Gated Photomultiplier Module User Manual

User Manual 14 pages · 586 KB · EN

The Licel PM-HV Gated Photomultiplier Module R9880U is a compact detector optimized for pulsed applications with high dynamic range capabilities. It combines a stabilized dynode chain for strong light pulses with fast rise times, allowing measurements across 5 orders of magnitude using both analog and photon counting methods.

The document provides detailed instructions on handling the PMT to prevent vacuum damage from helium or excessive light, operating safety regarding high voltage hazards, and configuring gating parameters. It also includes specific guidance for using the module in LIDAR systems, including signal limits and cathode structure considerations.

Read online

Licel R9880U specifications

Max. continuous output signal-6mV
Recommended HV setting (peak < 20mV)1000V
Max. allowed anode average current100uA over 30 seconds
Max. allowed average anode current (LIDAR)5mV in analog mode at a 50 Ω input averaged over a 100 ms time
Peak signal limit (100mV input range)50mV

Source: Licel R9880U Photomultiplier Module User Manual

All Licel R9880U documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the recommended high voltage setting if the peak signal amplitude is always below 20mV?

The recommended HV setting is 1000V.

From: Licel R9880U Photomultiplier Module User Manual

How should photomultipliers be stored to protect them?

They should be stored in the dark and protected from excessive light, especially during installation.

From: Licel R9880U Photomultiplier Module User Manual

What are the requirements for an external timing generator?

The generator must be able to deliver more than 2.5V into 50Ω.

From: Licel R9880U Gated Photomultiplier Module User Manual

What happens if a PMT is stored in a helium rich atmosphere?

Helium will diffuse through the housing and destroy the internal vacuum.

From: Licel R9880U Gated Photomultiplier Module User Manual