This manual introduces the features, usage methods and precautions of Licel PM-HV Gated Photomultiplier Module R9880U.
The Licel PM-HV Gated Photomultiplier Module R9880U is a compact detector optimized for pulsed applications with high dynamic range capabilities. It combines a stabilized dynode chain for strong light pulses with fast rise times, allowing measurements across 5 orders of magnitude using both analog and photon counting methods.
The document provides detailed instructions on handling the PMT to prevent vacuum damage from helium or excessive light, operating safety regarding high voltage hazards, and configuring gating parameters. It also includes specific guidance for using the module in LIDAR systems, including signal limits and cathode structure considerations.
| Maximum ON time (nominal gain) | 1 ms |
|---|---|
| Maximum duty cycle | 1% |
| Minimum gating pulse voltage | >2.5V |
| Maximum average anode current | 100uA over 30 seconds |
| Recommended HV setting for peak signal <20mV | 1000V |
The generator must be able to deliver more than 2.5V into 50Ω.
Helium will diffuse through the housing and destroy the internal vacuum.
The limit is 100µA averaged over a 30 seconds period.