This manual provides detailed information on the Licel PM-HV photomultiplier module R9880U, including product description, handling, operation, and using in a LIDAR.
The document provides a comprehensive guide for the Licel PM-HV Photomultiplier Module R9880U. It details product descriptions, proper handling procedures to prevent vacuum damage from helium or light exposure, and safety warnings regarding high voltage shock hazards.
Key operational topics include setting pulse height distributions, configuring discriminator thresholds, and using the device in LIDAR systems. The manual explains the differences between analog and photon counting modes, including dead time corrections for paralyzable and nonparalyzable systems.
| Max. continuous output signal | -6mV |
|---|---|
| Recommended HV setting (peak < 20mV) | 1000V |
| Max. allowed anode average current | 100uA over 30 seconds |
| Max. allowed average anode current (LIDAR) | 5mV in analog mode at a 50 Ω input averaged over a 100 ms time |
| Peak signal limit (100mV input range) | 50mV |
The recommended HV setting is 1000V.
They should be stored in the dark and protected from excessive light, especially during installation.
Helium will diffuse through the housing and destroy the internal vacuum.
The Licel photon counter can be best described as nonparalyzable.