JEOL JEM-2100 Electron Microscope Instruction Manual
The JEOL JEM-2100 Electron Microscope Instruction Manual is provided for proper use of the instrument; JEOL asks users to read it before use and keep it on hand for reference. It includes safety precautions and notices on installation, modification, maintenance, relocation, and servicing by JEOL.
The manual covers specifications, composition and construction, control panels, and PC windows, followed by operating procedures for startup and shutdown, film loading, specimen exchange, electron beam generation, aperture and lens adjustment, electron diffraction, special observation, the anti-contamination device, and photography.
JEOL JEM-2100 specifications
| ION PUMP vacuum condition | <2.5x10-5 Pa with blue scale |
|---|---|
| Column pressure before observation | below 1.5x10-5 Pa |
| TEM magnification for digital imaging | 40,000x |
| GIF image magnification | 40,000x |
| Digital camera temperature control | COOL, < -25oC |
| Spot size and alpha | Spot size = 1, alpha = 3 |
Source: JEOL JEM-2100F Transmission Electron Microscope Daily Operation Guide
All JEOL JEM-2100 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What should I do before using the JEM-2100 electron microscope?
Read the instruction manual before use and keep it on hand so you can consult it whenever necessary.
Who manufactures the JEM-2100?
JEOL Ltd. is listed as the manufacturer.
What should I do in an emergency?
Close the gun valve by pressing the “Beam” button at the upper-left corner of the left control panel.
From: JEOL JEM-2100F Transmission Electron Microscope Daily Operation Guide
What should the first user of the day do with the ACD?
Take out the ACD heater, fill the Anticontamination device with LN2, wait 20 minutes, and top off LN2 again.
From: JEOL JEM-2100F Transmission Electron Microscope Daily Operation Guide
How do I start the TEMCON software?
Double click the TEMCON icon after logging in with password 'JEOL' and waiting for the small circle to change from blue to red.
From: JEOL JEM-2100 Transmission Electron Microscope Operation Manual
What is the correct procedure to remove a sample?
Turn off the electron beam, set stage position to (x,0; y,0; z,0; t,0), pull sample holder from lens barrel, wait 30 sec, then insert sample into port and fit guide pin.
From: JEOL JEM-2100 Transmission Electron Microscope Operation Manual