The JENCO 6250/6250C is a portable multi-function meter designed for laboratory and field use. It measures pH, millivolts (mV), ion concentration, and temperature. The device features a large LCD display capable of showing multiple parameters simultaneously and supports both automatic temperature compensation (ATC) and manual temperature compensation (MAN). It can be powered by AAA batteries or an AC adapter. This manual is written in Chinese.

Read online

JENCO 6250/6250C specifications

pH range-2.000 to 16.000 pH
pH accuracy+/- 0.1 % +/- 1 digit
pH resolution0.01 pH or 0.001 pH, selectable
mV range-1999.9 to 1999.9 mV
mV resolution1 mV or 0.1 mV, selectable
Temperature range-5.0 to 125.0 °C
Ion range0.00 to 119.8 conc.
Dimensions241 mm / 83 mm / 72 mm (L/W/H)

Source: JENCO 6250/6250C Microcomputer pH/mV/Ion/Temp Portable Measuring Instrument Operation Manual

All JENCO 6250/6250C documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What parameters does the JENCO 6250/6250C meter measure?

It measures pH, mV, ion concentration, and temperature.

From: JENCO 6250/6250C Multi-function Meter User Manual

What is the pH measurement range?

The pH range is -2.000 ~ 16.000 pH.

From: JENCO 6250/6250C Multi-function Meter User Manual

What pH calibration solutions does the MODEL 6250 use?

MODEL 6250 uses 4.00 pH, 7.00 pH, and 10.01 pH standard calibration solutions.

From: JENCO 6250/6250C Microcomputer pH/mV/Ion/Temp Portable Measuring Instrument Operation Manual

What pH calibration solutions does the MODEL 6250C use?

MODEL 6250C uses 4.00 pH, 6.86 pH, and 9.18 pH standard calibration solutions.