Interscan 4000 Series Digital Portable Analyzer Instruction Manual

Interscan 4000 Series Digital Portable Analyzer Instruction Manual

User Manual 29 pages · 1 MB · EN

This document serves as the instruction manual for the Interscan 4000 Series Digital Portable Analyzer. It details the equipment description, including front, rear, and internal components, as well as operating instructions for setting alarms, zeroing the instrument, and sampling. The manual also provides guidance on calibration procedures using sample bags or pressurized cylinders.

Read online

Interscan 4000 Series Digital Portable Analyzer specifications

Battery Type ANickel - Cadmium (NiCd)
Battery Type BC size alkaline batteries
Nominal Sample Rate1.0, 0.2 liter per minute
Average Sample Run Time12 hours
Charger Input9V DC, 100mA

Source: Interscan 4000 Series Digital Portable Analyzer Instruction Manual

All Interscan 4000 Series Digital Portable Analyzer documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

Is calibration required when the analyzer is received?

No, it is not necessary to calibrate the monitor when received from the Interscan or an authorized distributor as they are calibrated at the factory.

From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual

What is the minimum alarm level setting?

The minimum alarm level must be greater than 5% of the full scale measuring range.

From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual

How often should batteries be checked?

Batteries are to be checked every 30 days.

From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual

What is the average sample run time?

The average sample run time is 12 hours with fully charged C NiCd batteries.

From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual