Instruction manual for the Interscan 4000 Series Digital Portable Analyzer covering operation, calibration, maintenance, and troubleshooting.
This document serves as the instruction manual for the Interscan 4000 Series Digital Portable Analyzer. It details the equipment description, including front, rear, and internal components, as well as operating instructions for setting alarms, zeroing the instrument, and sampling. The manual also provides guidance on calibration procedures using sample bags or pressurized cylinders.
| Battery Type A | Nickel - Cadmium (NiCd) |
|---|---|
| Battery Type B | C size alkaline batteries |
| Nominal Sample Rate | 1.0, 0.2 liter per minute |
| Average Sample Run Time | 12 hours |
| Charger Input | 9V DC, 100mA |
No, it is not necessary to calibrate the monitor when received from the Interscan or an authorized distributor as they are calibrated at the factory.
The minimum alarm level must be greater than 5% of the full scale measuring range.
You must allow 24 hours to stabilize prior to use if the batteries are replaced.