Interscan 4000 Series Digital Portable Analyzer Instruction Manual
This document serves as the instruction manual for the Interscan 4000 Series Digital Portable Analyzer. It details the equipment description, including front, rear, and internal components, as well as operating instructions for setting alarms, zeroing the instrument, and sampling. The manual also provides guidance on calibration procedures using sample bags or pressurized cylinders.
Read onlineINTERSCAN 4000 Series specifications
| Accuracy | ±2% Rd±1 unit minimum significant digit |
|---|---|
| Repeatability | ±0.5% F.S. |
| Minimum Detection Limit | ±1% F.S. |
| Linearity | ±1% F.S. |
| Zero Drift (24h) | ±1% F.S. |
| Span Drift (24h) | ±1% F.S. |
| Delay Time | < 1 second |
| Dimensions | 178mm×102mm×225mm |
Source: INTERSCAN 4000 Series Single Gas Analyzer Selection Guide
All INTERSCAN 4000 Series documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Do the analyzers require recalibration after purchase?
No, all INTERSCAN analyzers are calibrated before leaving the factory and do not need re-calibration unless the SPAN potentiometer is accidentally changed or the instrument is used for a long time.
From: INTERSCAN 4000 Series Single Gas Analyzer Selection Guide
What is the physical volume of the analyzer?
The physical volume of the analyzer is 178mm x 102mm x 225mm.
From: INTERSCAN 4000 Series Single Gas Analyzer Selection Guide
Is calibration required when the analyzer is received?
No, it is not necessary to calibrate the monitor when received from the Interscan or an authorized distributor as they are calibrated at the factory.
From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual
What is the minimum alarm level setting?
The minimum alarm level must be greater than 5% of the full scale measuring range.
From: Interscan 4000 Series Digital Portable Analyzer Instruction Manual
What is the nominal sample rate of the analyzer?
The nominal sample rate is approximately 1.0 liter per minute, ± 0.2 liter per minute.
From: Interscan 4000 Series Compact Portable Analyzer Instruction Manual
How long is the average sample time with fully charged “C” batteries?
The average sample time is 12 hours with fully charged “C” batteries.
From: Interscan 4000 Series Compact Portable Analyzer Instruction Manual