Key facts
Start here

Humboldt HM-4150 Distribution Panel Constant Head Permeability Quick Start Guide
This quick start guide for the Humboldt HM-4150 Distribution Panel describes the constant head permeability setup. It covers checking the base and top burette cross-sectional area, filling the top burette when needed, and connecting tubing to the top burette assembly and triaxial cell.
The guide then lists valve and tubing steps, including flushing the triaxial cell top input valve, closing all triaxial cell valves, and setting the base pressure and top pressure burette valves. It also includes a reference to instructions for performing falling head, rising tail permeability.
Humboldt HM-4150 specifications
| Document Model | HM-4150 |
|---|---|
| Document Type | Quick Start Guide |
| Copyright Year | 2018 |
Source: Humboldt HM-4150 Falling Head Permeability Distribution Panel Quick Start Guide
All Humboldt HM-4150 documents by type
The shortest path to a working setup, usually only a few pages.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
How is the area of the burette calculated?
The area of the burette equals the known volume divided by the measured length.
From: Humboldt HM-4150 Distribution Panel Constant Head Permeability Quick Start Guide
Which valve position is used when flushing the Top Input valve?
Turn the Top Pressure burette assembly’s Fill/Drain valve to the Fill position and allow the water to drain.
From: Humboldt HM-4150 Distribution Panel Constant Head Permeability Quick Start Guide
How do you verify the burette cross-sectional area?
Measure the length of a known volume of the burette using a steel ruler and divide the volume by the length.
From: Humboldt HM-4150 Falling Head Permeability Distribution Panel Quick Start Guide
What is the formula for calculating permeability?
Permeability K is calculated using the areas of the reservoirs, the length of the specimen, the average head loss, and the time interval.
From: Humboldt HM-4150 Falling Head Permeability Distribution Panel Quick Start Guide