Hart Scientific 5627A Manuals

ManualLib has 2 documents for the Hart Scientific 5627A (Temperature sensors): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
5627A
Industry
Industrial
Documents
2
Languages
EN
Last added
2026-11-16

Start here

Hart Scientific 5627A Platinum Resistance Thermometer User Guide

Hart Scientific 5627A Platinum Resistance Thermometer User Guide

User Manual 14 pages · 112 KB · EN

This guide introduces the 5627A Platinum Resistance Thermometer, a secondary‑standard instrument for precise temperature measurement. It covers general product information, recommended recalibration intervals, and detailed specifications such as resistance, temperature range, and construction materials.

The manual also includes safety warnings, lead‑wire identification, handling and care procedures, installation instructions, and troubleshooting tips. Contact information for authorized service centers worldwide is provided for warranty support and repairs.

Read online

Hart Scientific 5627A specifications

ResistanceNominal 100 Ω
Temperature Coefficient0.00385 Ω/Ω/ °C nominal
Temperature Range–200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
Drift Rate±0.13 °C at 0 °C after 1000 hours at 400 °C
Sheath Material316 Stainless Steel
LeadsTeflon™-insulated, nickel-plated stranded copper, 22 AWG
Time ConstantFour seconds maximum for 63.2 % response to step change in water moving at 3 fps
ImmersionAt least 100 mm (4 in) recommended

Source: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide

All Hart Scientific 5627A documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What can the 5627A be used as?

It may be used as a temperature standard for calibration of industrial sensors.

From: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide

How often is a PRT normally recalibrated?

A PRT is normally recalibrated annually.

From: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide