Hart Scientific 5627A Platinum Resistance Thermometer User Guide
This guide introduces the 5627A Platinum Resistance Thermometer, a secondary‑standard instrument for precise temperature measurement. It covers general product information, recommended recalibration intervals, and detailed specifications such as resistance, temperature range, and construction materials.
The manual also includes safety warnings, lead‑wire identification, handling and care procedures, installation instructions, and troubleshooting tips. Contact information for authorized service centers worldwide is provided for warranty support and repairs.
Hart Scientific 5627A specifications
| Resistance | Nominal 100 Ω |
|---|---|
| Temperature Coefficient | 0.00385 Ω/Ω/ °C nominal |
| Temperature Range | –200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C) |
| Drift Rate | ±0.13 °C at 0 °C after 1000 hours at 400 °C |
| Sheath Material | 316 Stainless Steel |
| Leads | Teflon™-insulated, nickel-plated stranded copper, 22 AWG |
| Time Constant | Four seconds maximum for 63.2 % response to step change in water moving at 3 fps |
| Immersion | At least 100 mm (4 in) recommended |
Source: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide
All Hart Scientific 5627A documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What can the 5627A be used as?
It may be used as a temperature standard for calibration of industrial sensors.
From: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide
How often is a PRT normally recalibrated?
A PRT is normally recalibrated annually.
From: Hart Scientific 5627A Platinum Resistance Thermometer User's Guide