The Hart Scientific 5627A Platinum Resistance Thermometer User Guide provides safety, installation, operation, care, and warranty information.
This guide introduces the 5627A Platinum Resistance Thermometer, a secondary‑standard instrument for precise temperature measurement. It covers general product information, recommended recalibration intervals, and detailed specifications such as resistance, temperature range, and construction materials.
The manual also includes safety warnings, lead‑wire identification, handling and care procedures, installation instructions, and troubleshooting tips. Contact information for authorized service centers worldwide is provided for warranty support and repairs.
| Resistance (Nominal) | 100 Ω |
|---|---|
| Temperature Coefficient | 0.00385 Ω/Ω/°C |
| Temperature Range | –200 °C to 420 °C |
| Drift Rate | ±0.13 °C at 0 °C after 1000 h at 400 °C |
| Sheath Material | 316 Stainless Steel |
| Lead Wire Type | Teflon‑insulated, nickel‑plated stranded copper, 22 AWG |
| Time Constant | 4 seconds (63.2 % response) |