User’s Guide for the Hart Scientific 5627A Platinum Resistance Thermometer, covering safety, specifications, installation, care, operation, and troubleshooting.
The Hart Scientific 5627A Platinum Resistance Thermometer User’s Guide covers safety information, specifications, installation, care and handling, operation, and troubleshooting for the 5627A PRT. The instrument is described as a secondary standard interpolating instrument that converts temperature to resistance and may also be used as a temperature standard for calibrating industrial sensors.
The guide explains the four-wire cable and lead wire identification, lists PRT care and handling guidelines, and gives specifications such as nominal 100 Ω resistance, a –200 °C to 420 °C temperature range, and a recommended immersion depth of at least 100 mm. It also addresses recalibration and authorized service centers.
| Resistance | Nominal 100 Ω |
|---|---|
| Temperature Coefficient | 0.00385 Ω/Ω/ °C nominal |
| Temperature Range | –200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C) |
| Drift Rate | ±0.13 °C at 0 °C after 1000 hours at 400 °C |
| Sheath Material | 316 Stainless Steel |
| Leads | Teflon™-insulated, nickel-plated stranded copper, 22 AWG |
| Time Constant | Four seconds maximum for 63.2 % response to step change in water moving at 3 fps |
| Immersion | At least 100 mm (4 in) recommended |
It may be used as a temperature standard for calibration of industrial sensors.
A PRT is normally recalibrated annually.
The 5627A PRT is equipped with a four-wire cable.
Hart Scientific recommends 1 mA.