Hart Scientific 5627A Platinum Resistance Thermometer User's Guide

User’s Guide for the Hart Scientific 5627A Platinum Resistance Thermometer, covering safety, specifications, installation, care, operation, and troubleshooting.


Hart Scientific 5627A Platinum Resistance Thermometer User's Guide - cover page
Brand
Hart Scientific
Category
Temperature sensors
Document type
User Manual
Language
English
Pages
12
File format
PDF
File size
124 KB
Published
16 November, 2026
Updated
16 November, 2026
Model
5627A
MD5 checksum
50F1CFABD1DB5010CE77DFB79816B66F

About this manual

The Hart Scientific 5627A Platinum Resistance Thermometer User’s Guide covers safety information, specifications, installation, care and handling, operation, and troubleshooting for the 5627A PRT. The instrument is described as a secondary standard interpolating instrument that converts temperature to resistance and may also be used as a temperature standard for calibrating industrial sensors.

The guide explains the four-wire cable and lead wire identification, lists PRT care and handling guidelines, and gives specifications such as nominal 100 Ω resistance, a –200 °C to 420 °C temperature range, and a recommended immersion depth of at least 100 mm. It also addresses recalibration and authorized service centers.

What's inside

  • Before You Start
  • Safety Information
  • Introduction
  • Recalibration
  • Specifications
  • Installation
  • Lead Wire Identification
  • PRT Care and Handling Guidelines
  • Operation
  • Troubleshooting

Specifications

ResistanceNominal 100 Ω
Temperature Coefficient0.00385 Ω/Ω/ °C nominal
Temperature Range–200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C)
Drift Rate±0.13 °C at 0 °C after 1000 hours at 400 °C
Sheath Material316 Stainless Steel
LeadsTeflon™-insulated, nickel-plated stranded copper, 22 AWG
Time ConstantFour seconds maximum for 63.2 % response to step change in water moving at 3 fps
ImmersionAt least 100 mm (4 in) recommended

Frequently asked questions

What can the 5627A be used as?

It may be used as a temperature standard for calibration of industrial sensors.

How often is a PRT normally recalibrated?

A PRT is normally recalibrated annually.

How many wires does the 5627A cable have?

The 5627A PRT is equipped with a four-wire cable.

What drive current does Hart Scientific recommend?

Hart Scientific recommends 1 mA.