EXFO FPM-300/FLS-300/FOT-300 Power Meter/Light Source/Optical Loss Test Set User Guide
This user guide provides instructions for operating the EXFO FPM-300 power meter, FLS-300 light source, and FOT-300 optical loss test set. It covers safety precautions, setup, measuring power or loss, using the light source, maintenance, troubleshooting, and warranty information. The devices use Ge or GeX detectors with 10 calibrated wavelengths and support automatic wavelength detection and shutdown after 10 minutes of inactivity.
The guide includes detailed sections on connecting optical fibers, nulling electrical offsets, defining favorite wavelengths, and recalibrating the unit. Safety information for laser use (FLS-300 and FOT-300) and electrical safety is provided. Technical specifications, compliance certifications, and contact information for EXFO service centers are also included.
EXFO FLS-300 specifications
| Size | 185 mm × 100 mm × 55 mm (7 ¼ in × 4 in × 2 1/8 in) |
|---|---|
| Weight | 0.4 kg (0.9 lb) |
| Operating temperature | -10 °C to 50 °C (14 °F to 122 °F) |
| Storage temperature | -40 °C to 70 °C (-40 °F to 158 °F) |
| Relative humidity | 0% to 95% non-condensing |
| Central wavelength, model 23BL | 1310 ± 20 nm and 1550 ± 20 nm |
| Central wavelength, model 12D | 850 ± 25 nm and 1300 +50/-20 nm |
| Spectral width, model 23BL | ≤5 nm |
| Spectral width, model 12D | 50/135 nm |
| Output power, model 23BL | ≥1/≥1 dBm |
| Output power, model 12D | ≥-20/≥-20 dBm (62.5/125 µm) |
| Power stability | ±0.10 dB (8 hours) |
| Battery life | 120 hours |
| Tone generation | 270, 1 k, 2 k Hz |
| Warranty and recommended calibration interval | 3 years |
All EXFO FLS-300 documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
Which wavelengths does model 23BL provide?
1310 nm and 1550 nm.
What are the three products covered by this user guide?
FPM-300 power meter, FLS-300 light source, and FOT-300 OLTS.
How do I power the device?
The device can be powered by an AC adapter (for indoor use) or AA alkaline batteries (which automatically take over if the AC adapter is unplugged).
From: EXFO FPM-300/FLS-300/FOT-300 Power Meter/Light Source/Optical Loss Test Set User Guide
What are the typical applications of this equipment?
Typical applications include transmitter power measurements (dBm and W), fiber-link loss testing (dB), component insertion-loss testing (dB), fiber identification with 270-Hz, 1-kHz and 2-kHz signals, fiber installation and maintenance, and FTTx testing of passive optical networks (PONs).