Key facts
- Brand
- Envirologix
- Model
- AQ-304-BG
- Category
- Other
- Documents
- 2
- Languages
- ES · EN
- Last added
- 2023-10-17
Start here

Envirologix AQ-304-BG QuickScan Test Kit Quick Start Guide
This document provides a quick start guide for the Envirologix AQ-304-BG QuickScan test kit. It outlines the necessary sample preparation steps for corn and wheat, including grinding, weighing, and extraction using distilled water and a coffee filter. The guide specifies an incubation temperature of 22°C and a sample weight of 20-50 grams. This manual is written in Spanish.
Read onlineEnvirologix AQ-304-BG specifications
| Incubation Temperature | 22°C |
|---|---|
| Sample Weight | 20-50 grams |
| Extraction Volume | 5 mL/g |
| Buffer Volume | 100 ul |
| Incubation Time | 2 minutes |
| Test Duration | 2 minutes |
| Measurement Range | 0 to 8 ppm |
Source: Envirologix AQ-304-BG QuickScan Test Kit Quick Start Guide
All Envirologix AQ-304-BG documents by type
The shortest path to a working setup, usually only a few pages.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What crops does this test kit support?
The kit is designed for corn and wheat samples.
From: Envirologix AQ-304-BG QuickScan Test Kit Quick Start Guide
What is the required incubation temperature?
The incubator must be configured to 22°C.
From: Envirologix AQ-304-BG QuickScan Test Kit Quick Start Guide
What temperature range requires acclimation of the sample strips?
Acclimation is only required when the ambient temperature is unknown or outside of 20-24°C (68-75°F).
What is the maximum concentration reported by the test results?
Results are reported between 0 and 8 ppm.