Envirologix AQ-314-BG T-2 Manuals

ManualLib has 2 documents for the Envirologix AQ-314-BG T-2 (Other): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
AQ-314-BG T-2
Category
Other
Documents
2
Languages
EN
Last added
2023-10-17

Start here

Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan User Manual

Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan User Manual

User Manual 1 pages · 1 MB · EN

This document provides instructions for using the Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan device. It details the sample preparation process, including grinding and sieving, as well as the specific steps for conducting the test using the QuickScan system. The device reports results ranging from 50 to 900 ppb.

Read online

Envirologix AQ-314-BG T-2 specifications

Reported Range50 - 900 ppb
Incubator Temperature22°C
Acclimation Temperature Range20-24°C (68-75°F)
Sample Sieve Size20 mesh

Source: Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan User Manual

All Envirologix AQ-314-BG T-2 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the reported range of the QuickScan device?

The device reports results from 50 to 900 ppb.

From: Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan User Manual

At what temperature should the incubator be set?

The incubator should be set to 22°C.

From: Envirologix AQ-314-BG T-2/HT-2 Flex QuickScan User Manual

What is the reported range for Extended Dilution results?

Extended Dilution results are reported between 900 and 2500 ppb.

From: Envirologix AQ-314-BG T-2/HT-2 Flex Extended Dilution Test Procedure

What is the correct dilution ratio for the extract?

You must dilute the extract with water at a 1:6 ratio.

From: Envirologix AQ-314-BG T-2/HT-2 Flex Extended Dilution Test Procedure