Dafang Huizhong LCZ-803 Manuals

ManualLib has 2 documents for the Dafang Huizhong LCZ-803 (Other products): 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Model
LCZ-803
Documents
2
Languages
ZH
Last added
2023-11-25

Start here

Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual

Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual

User Manual 18 pages · 3.9 MB · ZH

The LCZ-803 Series ultrasonic flowmeter is made by Tangshan Dafang Huizhong Instrument Co., Ltd. It measures liquid flow in circular pipes using the velocity difference method and uses multipulse technology, digital signal processing, and error correction.

The manual covers the flowmeter introduction, converter installation and connection, converter display and operation, sensor installation and use, cable laying, common fault diagnosis, and RS485 and Modbus communication appendices. It also lists technical specifications and selection codes. This manual is written in Chinese.

Read online

Dafang Huizhong LCZ-803 specifications

Measurement accuracyClass 1.0 (plug-in) / Class 1.5 (clamp-on)
Repeatability±0.5% (plug-in) / ±0.8% (clamp-on)
Flow velocity range±0.3 m/s to ±12.0 m/s
Pipe diameter range40–2000 mm (plug-in) / 50–2000 mm (clamp-on)
Medium temperature0°C to 150°C
Current outputOptically isolated 4–20 mA, max 500 m, load < 600 Ω
Digital communicationRS485, 2400/4800/9600 bps, max 1200 m
Power supplyAC 127 V (95–140 V), 50 Hz, ≤10 W
Protection ratingIP54
Explosion-proof typeMining flameproof and intrinsically safe, Exd[ib]I
Sensor cable lengthStandard 10 m, up to 200 m on order
Display2×16 graphic backlit LCD

Source: Dafang Huizhong LCZ-803 Mining Explosion-Proof Digital Ultrasonic Flowmeter Installation and Operation Manual

All Dafang Huizhong LCZ-803 documents by type

Questions answered in these manuals

What media can the LCZ-803 measure?

It measures homogeneous fluids filling the pipe with suspended solids below 10 g/L and particle size below 1 mm, such as tap water, clean water, sewage and industrial wastewater.

From: Dafang Huizhong LCZ-803 Mining Explosion-Proof Digital Ultrasonic Flowmeter Installation and Operation Manual

What sensor types are available for the LCZ-803?

Two types are available: plug-in sensors for pipe pressure up to 2 MPa, and clamp-on sensors mounted on the pipe outer wall, which have no pressure requirement and suit high-pressure or strongly corrosive fluids.

From: Dafang Huizhong LCZ-803 Mining Explosion-Proof Digital Ultrasonic Flowmeter Installation and Operation Manual

What measurement principle does the LCZ-803 Series use?

It uses the velocity difference method, based on the difference between ultrasonic propagation times in the upstream and downstream directions.

From: Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual

What liquids can the LCZ-803 Series measure?

It measures homogeneous fluids that fill the pipe, with suspended solids below 10 g/L and particle size below 1 mm, such as tap water, clean water, sewage, and industrial wastewater.

From: Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual