LCZ-803 Mining Explosion-proof and Intrinsically Safe Digital Ultrasonic Flowmeter is an instrument for measuring the flow rate of liquid in a circular pipe based on the principle of “velocity difference method”.
This manual covers the installation and use of the LCZ-803, a mining explosion-proof and intrinsically safe digital ultrasonic flowmeter made by Tangshan Dafang Huizhong Instrument Co., Ltd. The meter works on the velocity-difference principle and measures liquid flow in circular pipes. Its design meets the GB3836.1-2000, GB3836.2-2000 and GB3836.4-2000 requirements for electrical equipment in explosive gas atmospheres, and it carries the "MA" certification with safety mark number MFB020013.
The manual explains the flowmeter's working principle and technical specifications, installation and wiring of the converter, the converter display and key operation, installation of the plug-in and clamp-on sensors, cable laying, and maintenance notes with common fault diagnosis. Appendices describe the instrument status codes, the RS485 serial communication protocol and the Modbus communication protocol. This manual is written in Chinese.
| Measurement accuracy | Class 1.0 (plug-in) / Class 1.5 (clamp-on) |
|---|---|
| Repeatability | ±0.5% (plug-in) / ±0.8% (clamp-on) |
| Flow velocity range | ±0.3 m/s to ±12.0 m/s |
| Pipe diameter range | 40–2000 mm (plug-in) / 50–2000 mm (clamp-on) |
| Medium temperature | 0°C to 150°C |
| Current output | Optically isolated 4–20 mA, max 500 m, load < 600 Ω |
| Digital communication | RS485, 2400/4800/9600 bps, max 1200 m |
| Power supply | AC 127 V (95–140 V), 50 Hz, ≤10 W |
| Protection rating | IP54 |
| Explosion-proof type | Mining flameproof and intrinsically safe, Exd[ib]I |
| Sensor cable length | Standard 10 m, up to 200 m on order |
| Display | 2×16 graphic backlit LCD |
It measures homogeneous fluids filling the pipe with suspended solids below 10 g/L and particle size below 1 mm, such as tap water, clean water, sewage and industrial wastewater.
Two types are available: plug-in sensors for pipe pressure up to 2 MPa, and clamp-on sensors mounted on the pipe outer wall, which have no pressure requirement and suit high-pressure or strongly corrosive fluids.
The converter uses AC 127 V (allowed range 95–140 V) at 50 Hz, or DC 12–36 V, consuming no more than 10 W.