Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual

LCZ-803 series ultrasonic flowmeter adopts advanced multi-pulse technology, signal digitization processing technology and error correction technology, and adopts advanced ultrasonic flow sensor, which has the characteristics of high flow measurement accuracy, good stability, strong anti-interference ability and good fluid adaptability.


Dafang Huizhong LCZ-803 Series Ultrasonic Flowmeter Installation and Instruction Manual - cover page
Brand
Dafang Huizhong
Category
flow transmitter
Document type
User Manual
Language
Chinese
Pages
18
File format
PDF
File size
3.9 MB
Published
11 November, 2023
Updated
11 November, 2023
MD5 checksum
B5FFC8FC97CF78D7E509C9696A559183

About this manual

The LCZ-803 Series ultrasonic flowmeter is made by Tangshan Dafang Huizhong Instrument Co., Ltd. It measures liquid flow in circular pipes using the velocity difference method and uses multipulse technology, digital signal processing, and error correction.

The manual covers the flowmeter introduction, converter installation and connection, converter display and operation, sensor installation and use, cable laying, common fault diagnosis, and RS485 and Modbus communication appendices. It also lists technical specifications and selection codes. This manual is written in Chinese.

What's inside

  • Preface
  • Flowmeter Introduction
  • Converter Installation and Connection
  • Converter Display and Operation Instructions
  • Sensor Installation and Use Instructions
  • Cable Laying
  • Common Fault Diagnosis
  • Instrument Working Status Code Description
  • RS485 Serial Communication Protocol
  • Modbus Communication Protocol

Specifications

ModelLCZ-803
Measurement accuracy1.0 class (repeatability 0.5%); 1.5 class (repeatability 0.8%)
Flow velocity range±0.25 m/s to ±12.0 m/s
Pipe diameter range40 mm to 2000 mm; 50 mm to 2000 mm
Power supplyAC 220 V ±15%, 50 Hz; AC 110 V ±10%, 60 Hz (optional); DC 12 V to 36 V, 1 A (optional)
Digital communicationRS485 with adjustable 2400, 4800, or 9600 bps; maximum transmission distance 1200 m
Protection classWall-mounted converter: IP65; panel-mounted converter: IP51; sensor: IP68
Signal cable lengthStandard 10 m; optional, maximum 200 m

Frequently asked questions

What measurement principle does the LCZ-803 Series use?

It uses the velocity difference method, based on the difference between ultrasonic propagation times in the upstream and downstream directions.

What liquids can the LCZ-803 Series measure?

It measures homogeneous fluids that fill the pipe, with suspended solids below 10 g/L and particle size below 1 mm, such as tap water, clean water, sewage, and industrial wastewater.

What is the standard signal cable length?

The standard length is 10 m, and lengths up to 200 m can be ordered.

Which communication protocols are supported?

The flowmeter supports RS485 communication with Dafang custom protocol and Modbus RTU.