Bühler GAS 222 Manuals

ManualLib has 2 documents for the Bühler GAS 222 (Industrial Supply): 2 product leaflets. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 2

Key facts

Model
GAS 222
Documents
2
Languages
DE
Last added
2012-07-10

Start here

Buhler Technologies GAS 222.21 Gas Extraction Probe Technology Data

Buhler Technologies GAS 222.21 Gas Extraction Probe Technology Data

Product Leaflet 2 pages · 403 KB · DE

This document provides technical specifications for the Buhler Technologies GAS 222.21 gas extraction probe (Sonde). The device is designed with a material of 1.4571 and features a ball valve made of 1.4408, utilizing PTFE/Graphite seals. This manual is written in German.

Read online

Bühler GAS 222 specifications

Probe Material1.4571
Valve Material1.4408
Seals MaterialPTFE/Graphite
Max Operating Temperature (Probe)200°C
Max Operating Pressure6 bar
Supply Voltage (Option A)230 V, 2.0A, 50/60Hz
Supply Voltage (Option B)115 V, 3.8 A, 50/60Hz
Temperature Range (Controller)+50 to +200°C
Alarm Setting Range5...30K from setpoint
Switching Current Max1A
Protection ClassIP54
Ambient Temperature Range-20 to +70°C

Source: Buhler Technologies GAS 222.21 Gas Extraction Probe Technology Data

All Bühler GAS 222 documents by type

Questions answered in these manuals

What is the part number for the 230V version?

The part number for the 230V version is Art.Nr. 4622221.

From: Buhler Technologies GAS 222.21 Gas Extraction Probe Technology Data

What is the part number for the 115V version?

The part number for the 115V version is Art.Nr. 4622225.

From: Buhler Technologies GAS 222.21 Gas Extraction Probe Technology Data

What is the maximum temperature rating for Inconel extraction tubes?

Inconel extraction tubes are rated for a maximum temperature of 1050°C.

From: Bühler Technologies GAS 222 Technology Data

What voltage is required to operate the heated extension GF?

The heated extension GF operates at 230V.

From: Bühler Technologies GAS 222 Technology Data