User manual for the Yitiancheng GYD Series dry type test transformer, covering operation, safety, and specifications for insulation testing.
This manual provides instructions for the Yitiancheng GYD Series dry type test transformer, a lightweight and compact device designed for on-site insulation performance testing of electrical equipment. It details the product's introduction, general technical indicators, working principles, and specific usage methods. This manual is written in Chinese.
| Output Voltage Waveform | Sine wave |
|---|---|
| Output Waveform Distortion Rate | ≤1% |
| No-load Loss | 0.2%~0.35% |
| Induced Withstand Voltage Level | 1.1U O/1min |
| Short-circuit Impedance | 4.5%~12%(±10%) |
| Voltage Adjustable Accuracy | ±1% |
| Voltage Instability | Less than 1% |
| Surface Temperature Rise | <55℃ |
| Continuous Operation Time (Rated Capacity) | 0.5 hours |
| Operating Temperature Range | 0—40℃ |
| Maximum Altitude | <1000m |
| Relative Humidity | <85% |
The product is covered by a free repair warranty for three years from the date of shipment.
No, the manual explicitly states not to operate the product in a humid environment.
The transformer allows for 0.5 hours of continuous operation at rated capacity.