User manual for the Yinhe Dianqi VSV301-G150T03-I voltage sensor, covering electrical and accuracy specifications, terminals, connections, and precautions.
The VSV301-G150T03-I is a voltage sensor from Yinhe Dianqi that measures DC, AC, pulse, and irregular waveforms with complete isolation between the primary and secondary sides. It is intended for metrology verification and calibration, laboratory voltage measurement, power analyzers, medical equipment such as MRI, battery pack testing, power control, power supplies, ships, new energy, rail transit, aerospace, and industrial measurement.
This user manual provides installation, operation, and maintenance notes. It describes the sensor's technical features, electrical performance, accuracy and dynamic parameters, general and safety characteristics, dimensions, terminal definitions, electrical connection diagram, measurement formula, packing list, and precautions. This manual is written in Chinese.
| Primary rated voltage (VPN) | 300 Vrms |
|---|---|
| Voltage measurement range (VPM) | 500 Vp |
| Supply voltage (Vc) | ±15 Vdc |
| Current consumption (Ic) | ±80 mA |
| Turns ratio (KN) | 500:10 |
| Secondary output (VS) | ±10 Vp |
| Measurement load (RM) | 5 kΩ |
| Accuracy (Xe) | 0.2 % |
| Linearity (εL) | 100 ppm |
| Phase displacement (Δφ) | 0.0315° at 50 Hz |
| Frequency bandwidth (-3 dB) | 0 to 200 kHz |
| Mass (m) | 350 ± 30 g |
The primary rated voltage is 300 Vrms.
Using VP = KN × Us, where KN is the turns ratio and Us is the voltage drop across the measurement output.
The sensor has a 6 kV DC withstand voltage between the primary and secondary sides and between the secondary side and the case.