Operating conditions and specifications for the Tosoh Bioscience TSKgel G3000-HHR GPC column, covering flow rates, pressure, solvents and QC specs.
This two-page data sheet from Tosoh Bioscience lists the recommended operating conditions and specifications for TSKgel G3000-HHR columns and their guard columns. It covers the shipping solvent, maximum and standard flow rates, maximum pressure, multiple-column connection, compatible solvents, temperature range, sample size, storage and column protection. H-type columns are stated to be used exclusively for Gel Permeation Chromatography, and installation and column care are described in a separate instruction manual.
The second page gives the quality control specifications every column has passed, namely the number of theoretical plates and the asymmetry factor, tested under the conditions described in the data sheet. The listed products are part number 0017355 (7.8 mm ID x 30.0 cm L, G3000-HHR, 5 µm) and part number 0017368 (6.0 mm ID x 4.0 cm L, Guardcolumn HHR-L, 13 µm).
| Shipping solvent | Tetrahydrofuran (THF) |
|---|---|
| Maximum flow rate | 2.0 mL/min |
| Standard flow rate | 0.5 - 1.0 mL/min |
| Maximum pressure | 5 MPa |
| Temperature range | 25°C - 140°C |
| Sample size | 0.001 - 0.5 mg |
| Number of theoretical plates (N) | > 16,000 |
| Asymmetry factor (AF) | 0.7 - 1.6 |
| Analytical column part number | 0017355, 7.8 mm ID x 30.0 cm L, 5 µm |
| Guard column part number | 0017368, 6.0 mm ID x 4.0 cm L, 13 µm |
The column is shipped in tetrahydrofuran (THF).
The maximum flow rate is 2.0 mL/min and the maximum pressure is 5 MPa; with a high-viscosity buffer the flow rate may have to be reduced.
The listed compatible solvents include benzene, chloroform, toluene, dichloromethane, carbon tetrachloride, DMF, DMSO, dioxane, n-hexane, acetone, ethanol, methanol and pyridine.
Remove the column from the equipment, seal the ends with the provided protective screws and store it at laboratory temperature, keeping air out of the column.