Operating conditions and technical specifications for Tosoh Bioscience TSKgel DNA-NPR columns, including flow rates, pressure, and cleaning solvents.
This document provides the recommended operating conditions and specifications for TSKgel DNA-NPR columns. It details the necessary steps for initial flushing, buffer equilibration, and general column maintenance.
The sheet covers technical limits such as maximum flow rate, pressure, and pH range, while specifying quality control standards for theoretical plates and asymmetry factors.
| Max. Flow Rate | 1.5 mL/min |
|---|---|
| Standard Flow Rate | 0.5 - 1.0 mL/min |
| Max. Pressure | 30 MPa |
| pH Range | 2 - 12 |
| Temperature | 0 - 60 °C |
| Number of Theoretical Plates (N) | > 6,000 |
| Asymmetry Factor (AF) | 1.40 – 3.80 |
Flush the column and guard column with 30% acetonitrile/70% water for 15 minutes at 1 ml/min.
Recommended solvents include 0.1 - 0.2 M NaOH, 20 - 40% acetic acid aq., 30% acetonitrile or methanol in water or buffer, or urea/nonionic surfactant in buffer.
Store the column in water if it will not be used within the next three days, ensuring air does not enter the column.