This document lists operating conditions and technical specifications for TSKgel UP-SW3000 columns and guard columns, including flow rates, pressure limits…
This document provides the recommended operating conditions and specifications for TSKgel UP-SW3000 columns and guard columns. It details part numbers, recommended flow rates, maximum pressure limits, temperature ranges, and pH ranges. It also outlines procedures for cleaning the columns to remove basic substances, hydrophobic substances, and poorly soluble proteins, as well as proper storage methods.
| Column Length (Analytical) | 15.0 cm L or 30.0 cm L |
|---|---|
| Column Length (Guard) | 2.0 cm L |
| Particle Size | 2.0 µm |
| Standard Flow Rate | 0.10 - 0.35 mL/min |
| Max. Pressure (15 cm) | 25 MPa |
| Max. Pressure (30 cm) | 34 MPa |
| Temperature Range | 10 – 30 °C |
| pH Range | 2.5 – 7.5 |
| Organic Concentration | 0 - 30% |
| Number of Theoretical Plates (15 cm) | > 25,000 |
| Number of Theoretical Plates (30 cm) | > 45,000 |
| Asymmetry Factor | 0.90 – 1.40 |
The standard flow rate is between 0.10 and 0.35 mL/min.
The maximum pressure is 25 MPa for the 15 cm length and 34 MPa for the 30 cm length.
After washing the tubing, replace the column contents with the shipping solvent, seal both ends with end plugs, and store at 15 to 30°C.
The recommended pH range is 2.5 to 7.5.