Roche CoaguChek XS Pro System Operator's Manual

This manual provides operating instructions for the Roche CoaguChek XS Pro system, covering setup, testing procedures, quality control, and maintenance for…


Roche CoaguChek XS Pro System Operator's Manual - cover page
Brand
Roche
Category
Other
Document type
User Manual
Language
English
Pages
168
File format
PDF
File size
2.7 MB
Published
16 November, 2026
Updated
16 November, 2026
MD5 checksum
913E4231349FF1717D88B6B59F73C26B

About this manual

The CoaguChek XS Pro System Operator's Manual serves as a comprehensive guide for medical professionals using the CoaguChek XS Pro device. It details the intended use, safety instructions, and step-by-step procedures for operating the meter, including inserting batteries, powering the device, and performing blood tests on capillary or venous whole blood samples. The document also explains how to configure system settings, manage patient and operator lists, and handle data via barcode scanning and infrared interfaces.

The manual further covers essential maintenance tasks such as cleaning and disinfecting the meter using approved agents, as well as troubleshooting common status and error messages. It includes technical specifications regarding operating conditions, memory capacity, and power requirements, ensuring users can maintain the device correctly and safely in a clinical environment.

What's inside

  • Introduction
  • Safety Information
  • Overview of the Meter
  • Meter Setup
  • Testing a Blood Sample
  • Control Testing and Quality Control
  • Review Results
  • Extended Functionalities
  • Maintenance and Care
  • Troubleshooting

Specifications

Temperature range+15 °C to +32 °C (59 °F to 90 °F)
Relative humidity10 to 85% (no condensation)
Maximum altitude4300 m (14,000 feet)
Measuring range %Q120 – 5
Measuring range SEC9.6 – 96
Measuring range INR0.8 – 8.0
Memory capacity2000 patient and 500 QC results
Battery operation4 x 1.5 V batteries type AA
Number of tests per set of batteriesApprox. 80 tests
Dimensions231 x 97 x 43 mm
Weight350 g (without batteries)
Sample sizeAt least 8 μL

Frequently asked questions

What is the operating temperature range for the CoaguChek XS Pro?

The system should be operated within a temperature range of +15 °C to +32 °C (59 °F to 90 °F).

What type of batteries does the meter use?

The meter operates using 4 x 1.5 V batteries type AA or a special rechargeable battery pack.

How many test results can the memory store?

The memory can store up to 2000 patient results and 500 QC results with date and time.

What is the minimum sample size required for testing?

The system requires a sample size of at least 8 μL.