The Motic BA310E Biological Microscope Instruction Manual gives setup, operation, maintenance, and troubleshooting instructions for the BA310E microscope.
This instruction manual covers the Motic BA310E biological microscope, which is offered in binocular and trinocular configurations. It is based on infinity corrected optics, and the manufacturer notes that descriptions, illustrations, and specifications are subject to change without notice.
The manual provides operating environment limits, assembly instructions for components such as objectives, the condenser, eyepieces, and filters, and guidance on focusing, illumination, and brightness adjustment. It also covers photomicrographic procedures, oil immersion objectives, troubleshooting, care and maintenance, bulb replacement, and the warning labels found on the microscope.
| Altitude | Max 2000 meters |
|---|---|
| Ambient temperature | 15°C to 35°C |
| Supply voltage fluctuations | Not to exceed ±10% of the normal voltage |
| Air pressure | 75kPa to 106 kPa |
It uses infinity corrected optics, in which light beams are parallel after leaving the objective toward the eyepieces.
The ambient temperature should be between 15°C and 35°C.
Turn the power switch off and unplug the power cord, then firmly insert the LED module into the socket pinholes and secure it with the clamp screw using the supplied 1.5mm hexagonal screwdriver.