Mettler-Toledo ACT100xx weighing transmitter Zone 2 installation manual covering installation rules, special conditions, temperature class and wiring diagrams.
This is the Chinese-language Zone 2 installation manual for the Mettler-Toledo ACT100xx weighing transmitter (document 30529642 | 00 | 06/2020). It opens with Mettler-Toledo service information such as product registration, service agreements, initial calibration documentation, periodic calibration and GWP verification, followed by copyright, FCC and disposal notices, and a warnings section.
The main chapters cover the introduction and product marking, installation regulations and requirements, non-sparking certification, temperature class, special requirements for the housing and special conditions of use, and the certification documents including the installation wiring diagram. The manual states that installation in Zone 2 must refer to drawing 30562232 and that the device must not be used in Zones 0, 1, 20 or 21. This manual is written in Chinese.
| Temperature class (gas) | T4 (135 °C) |
|---|---|
| Permitted hazardous area | Zone 2 only; not for Zones 0, 1, 20, 21 |
| Installation drawing for Zone 2 | 30562232 |
| FCC compliance | FCC Rules Part 15, Class A digital device limits |
| Quality certifications of the producing factory | ISO9001, ISO14001, GB/T28001 |
The ACT100xx may be installed in Zone 2 only; it must not be used in flammable or explosive gas atmospheres of Zones 0, 1, 20 or 21.
For gases the transmitter has temperature class T4 (135 °C), so it must not be used where the ignition temperature of the hazardous substance is below this class.
Installation drawing 30562232 must be referred to; improper installation can cause personal injury and property damage.
The power supply to the device must always be disconnected before cleaning or performing maintenance.