Megger TDR1000/3 TDR1000/3P CFL510G TDR500/3 Time Domain Reflectometers User Manual

User manual for Megger's TDR1000/3, TDR1000/3P, CFL510G, and TDR500/3 time domain reflectometers, covering operation, safety, and specifications.


Megger TDR1000/3 TDR1000/3P CFL510G TDR500/3 Time Domain Reflectometers User Manual - cover page
Brand
Megger
Document type
User Manual
Language
Chinese
Pages
28
File format
PDF
File size
989 KB
Published
23 November, 2026
Updated
23 November, 2026
MD5 checksum
0387BBF4D791C3DAD57CEFFAB1FE6CF4

About this manual

This user manual provides detailed instructions for operating Megger’s TDR1000/3, TDR1000/3P, CFL510G, and TDR500/3 time domain reflectometers. It covers instrument layout, controls, testing procedures, safety precautions, and specifications. The device uses pulse echo (TDR) technology to locate cable faults by analyzing reflected pulses. Safety warnings emphasize de-energizing circuits and using correct test leads. The manual also explains how to adjust velocity factor, set impedance, and interpret reflections for fault detection.

The instrument supports various cable types (25 Ω, 50 Ω, 75 Ω, 100 Ω) and offers range settings from 10m to 5000m (30ft to 15000ft). It features a 256x128 dot matrix display, adjustable gain, and battery operation. Safety and EMC compliance are detailed, including EN 61010-1 for live systems and BS EN 61326-1 for EMC. Repair and warranty information is provided, including contact details for authorized service centers. This manual is written in Chinese.

What's inside

  • Introduction
  • Overview
  • Instrument Layout and Display
  • Controls
  • Operation
  • Safety Warnings
  • Specifications
  • Test Lead Selection
  • Repair and Warranty
  • Megger Contact Information

Specifications

General Ranges10 m, 25 m, 100 m, 250 m, 1000 m, 2500 m, 5000 m (30 ft, 75 ft, 300 ft, 750 ft, 3000 ft, 7500 ft, 15000 ft)
Accuracy±1% of range ± pixel at 0.67 VF
Resolution1% of range
Output Pulse5 volts nominal peak to peak into open circuit
Battery TypeFive LR6 (AA) type batteries, Manganese alkali or nickel metal-hydride cells
Battery Life14 hours typical
Velocity Factor RangeVariable from 0.2 to 0.99 in steps of 0.01
Power DownAutomatic after 5 minutes with no key press
BacklightStays on for 1 minute with no key press when activated

Frequently asked questions

What safety precautions should be taken when using the TDR?

The circuit under test must be switched off, de-energised, securely isolated and proved safe before test connections are made. Use the correct lead set, and ensure hands remain behind guards of probes/clips.

How do I adjust the velocity factor for accurate distance measurement?

Adjust the velocity factor to match that of the cable under test. If the VF is unknown, set the cursor to the end of the cable and adjust the VF until the correct cable length is displayed.

What types of cables can this instrument test?

It can test any cable with at least two separate conductive elements, including those with armouring or screen. It supports 25 Ω, 50 Ω, 75 Ω, and 100 Ω cables.

What is the battery life of the instrument?

Typical battery life is 14 hours.