French quick reference for the Life Technologies GlobalFiler Express PCR Amplification Kit covering PCR, electrophoresis requirements, and data analysis.
This quick reference is for the Life Technologies GlobalFiler Express PCR Amplification Kit and summarizes product information, prerequisites, PCR setup, allelic ladder requirements, electrophoresis instrument requirements, data analysis, and limited warranty.
It also provides thermal cycling conditions, amplified DNA storage guidance, reagent volumes for electrophoresis sample preparation, and instrument-specific electrophoresis conditions. This manual is written in French.
| Kit size | 200 reactions (Part no. 4476609) or 1000 reactions (Part no. 4474665) |
|---|---|
| Amplified loci | 21 autosomal STR loci, Amelogenin, 1 Y STR locus, and 1 Y indel locus |
| Approved thermal cyclers | Veriti Thermal Cycler or GeneAmp PCR System 9700 with gold-plated or silver sample block |
| Amplified DNA storage (<2 weeks) | 2 to 8 °C |
| Amplified DNA storage (>2 weeks) | -15 to -25 °C |
| GeneScan 600 LIZ Size Standard v2.0 volume | 0.5 µl per reaction |
| Hi-Di Formamide volume | 9.5 µl per reaction |
| Final extension | 60 °C for 8 min |
The kit is approved for use with the Veriti Thermal Cycler or the GeneAmp PCR System 9700 with a gold-plated or silver sample block.
Amplified DNA should be stored at 2 to 8 °C for less than 2 weeks or at -15 to -25 °C for more than 2 weeks.
Each reaction uses 0.5 µl of GeneScan 600 LIZ Size Standard v2.0 and 9.5 µl of Hi-Di Formamide.
For the 3130 instrument, one allelic ladder injection is recommended for every four injections.