Operating instructions for the LEYBOLD TURBOVAC TW 250 S turbomolecular pump and TURBO.DRIVE S frequency converter, covering installation, operation…
This document provides operating instructions for the LEYBOLD TURBOVAC TW 250 S turbomolecular pump and the TURBO.DRIVE S frequency converter. The pump is designed to evacuate vacuum chambers down to high vacuum pressure levels and is suitable for pumping air and clean gases. It requires a forevacuum pump and the TURBO.DRIVE S frequency converter for operation. The manual details the design, standard equipment, and ordering data for the system.
| High-vacuum connection | DN 100 ISO-K |
|---|---|
| Pumping speed for N2 | 230 l·s-1 |
| Ultimate pressure with two-stage, oil-sealed rotary vane pump | < 10 -9 mbar |
| Operating speed | 51 600 r.p.m. |
| Run-up time | approx. 3 min |
| Forevacuum connection | DN 16 KF |
| Weight | approx. 5 kg |
| Noise level | < 55 dB(A) |
It is suitable for pumping air and clean gases.
The maximum permissible forevacuum pressure is 3 mbar.
The supply voltage is 24 V DC ±5%.