PHOENIX L300i is a leak detector produced by Leybold. It can perform leak detection in vacuum, partial flow and sniff mode.
This document is the operating manual for the Leybold PHOENIX L300i leak detector. It provides instructions on safety precautions, design and function, and supplied equipment. The guide details various vacuum methods including Vacuum Diagram, Partial Flow Method, and Sniffer Mode.
The manual covers installation procedures such as placement, ambient conditions, electrical connections, and vacuum connections. It also explains the operation of the device, including start-up, display modes, control panel usage, interfaces, calibration routines (internal and external), and settings management.
In FINE mode, the minimum detectable leak rate is less than or equal to 5 · 10-12 mbar l / s.
The unit must run for at least 20 minutes with the gas ballast valve opened to free the pump oil from steam.
The device must be operated within a temperature range between +10°C and +40°C.