Integra manual surgical instruments product information and instructions covering intended use, warnings, cleaning, sterilization, and reprocessing.
This document from Integra LifeSciences covers manual surgical instruments used in multiple surgical specialties for basic tasks such as cutting, grasping, clamping, dissecting, probing, retracting, draining, aspirating, suturing, or ligating. It lists intended use, indications, patient-contacting materials, contraindications, and pre-use handling and inspection points.
The manual also gives warnings and precautions, reprocessing instructions for reusable instruments, cleaning and sterilization guidance, drying times for instrument trays, and storage guidance. It states that the instruments are supplied non-sterile and must be cleaned and sterilized before first use and any reuse.
| Patient-contacting materials | Ceramic, titanium nitride, stainless steel, tungsten, thermoset polymers (including silicone), thermoplastic polymers, or chrome plated brass |
|---|---|
| Drying time, instruments and aluminum/stainless steel trays | 15 to 30 minutes and visual inspection for dryness |
| Drying time, polymer-based instrument trays | Minimum 40 minutes and visual inspection for dryness |
| Supply condition | Provided non-sterile; must be cleaned and sterilized before first use and any reuse |
| Cleaning validation | Validated for effective cleaning using an automatic washer/disinfector cycle with pre-wash, enzyme wash, wash, and rinse phases |
| Wash 1 temperature | 66°C set point |
No. They are provided non-sterile and must be cleaned and sterilized before first use and any reuse.
No. The document states that these instruments have not been validated for flash sterilization.
They can consist of ceramic, titanium nitride, stainless steel, tungsten, thermoset polymers including silicone, thermoplastic polymers, or chrome plated brass.
Rinse the instrument immediately in distilled or demineralized water and keep it moist before decontamination, cleaning, and sterilization.