Integra Manual Surgical Instruments Product Information and Instructions

Integra manual surgical instruments product information and instructions covering intended use, warnings, cleaning, sterilization, and reprocessing.


Integra Manual Surgical Instruments Product Information and Instructions - cover page
Brand
Integra
Category
Professional medical equipment
Document type
User Manual
Language
English
Pages
40
File format
PDF
File size
1.1 MB
Published
08 November, 2026
Updated
08 November, 2026
MD5 checksum
1B267FE7DDF86D039C720BE3FA57E64D

About this manual

This document from Integra LifeSciences covers manual surgical instruments used in multiple surgical specialties for basic tasks such as cutting, grasping, clamping, dissecting, probing, retracting, draining, aspirating, suturing, or ligating. It lists intended use, indications, patient-contacting materials, contraindications, and pre-use handling and inspection points.

The manual also gives warnings and precautions, reprocessing instructions for reusable instruments, cleaning and sterilization guidance, drying times for instrument trays, and storage guidance. It states that the instruments are supplied non-sterile and must be cleaned and sterilized before first use and any reuse.

What's inside

  • Intended Use
  • Indications For Use
  • Materials
  • Contraindications
  • Pre-use/Handling/Inspection
  • Warnings
  • Precautions
  • Reprocessing Instructions for Reusable Instruments
  • Cleaning
  • Storage

Specifications

Patient-contacting materialsCeramic, titanium nitride, stainless steel, tungsten, thermoset polymers (including silicone), thermoplastic polymers, or chrome plated brass
Drying time, instruments and aluminum/stainless steel trays15 to 30 minutes and visual inspection for dryness
Drying time, polymer-based instrument traysMinimum 40 minutes and visual inspection for dryness
Supply conditionProvided non-sterile; must be cleaned and sterilized before first use and any reuse
Cleaning validationValidated for effective cleaning using an automatic washer/disinfector cycle with pre-wash, enzyme wash, wash, and rinse phases
Wash 1 temperature66°C set point

Frequently asked questions

Are the instruments sterile when supplied?

No. They are provided non-sterile and must be cleaned and sterilized before first use and any reuse.

Can these instruments be flash sterilized?

No. The document states that these instruments have not been validated for flash sterilization.

What materials can the instruments be made of?

They can consist of ceramic, titanium nitride, stainless steel, tungsten, thermoset polymers including silicone, thermoplastic polymers, or chrome plated brass.

What should be done immediately after use?

Rinse the instrument immediately in distilled or demineralized water and keep it moist before decontamination, cleaning, and sterilization.