EM Test NX30 Electrostatic Discharge Simulator

Datasheet for EM Test's NX30 ESD simulator, up to 30 kV, for automotive, medical, telecom, and aerospace ESD testing.


EM Test NX30 Electrostatic Discharge Simulator - cover page
Brand
emtest
Document type
Datasheet
Language
English
Pages
4
File format
PDF
File size
1.2 MB
Published
15 November, 2026
Updated
15 November, 2026
Model
NX30
MD5 checksum
20935204270CE5083DF98309A43C18AB

About this manual

The EM Test NX30 is an electrostatic discharge simulator capable of generating ESD pulses up to 30 kV in both air and contact discharge modes. It complies with standards such as IEC 61000-4-2 and ISO 10605, and is suitable for automotive, medical, broadcast, telecom, and aerospace applications. The unit includes a high-voltage base unit, discharge gun, interface, interlock, and self-calibration. It supports multiple discharge networks and offers options like vertical coupling planes and fiber optic links for advanced testing.

The device features a color touch display, adjustable discharge levels, and a built-in battery tested to 30,000 pulses at 30 kV. It operates on AC mains or DC supply and is designed for use in environments ranging from 5°C to 40°C with non-condensing humidity up to 80%.

What's inside

  • Test Voltage and Discharge Modes
  • Technical Specifications
  • Discharge Gun Controls
  • Accessories and Options
  • Environmental Specifications
  • Contact Information

Specifications

Test VoltageUp to 30 kV
Discharge ModesAir and Contact Discharge
Resolution100 V steps
Rise Time0.8 ns +/- 25%, 0.7 ns - 1 ns
Current Value at 30ns4.0 A @2 kV, 8.0 A @4 kV, 12.0 A @6 kV, 16.0 A @8 kV
Operating Temperature5°C to 40°C
Humidity20% to 80%, non condensing
Supply VoltageAC: 100 V - 240 V, 50/60 Hz
Weight (Base Station)Approx. 6.5 kg
Weight (Discharge Gun)Approx. 1,250 g

Frequently asked questions

What is the maximum test voltage of the NX30?

The NX30 can generate ESD pulses up to 30 kV in both air and contact discharge modes.

What discharge modes are supported?

The NX30 supports both air discharge and contact discharge modes.

What is the operating temperature range?

The device operates between 5°C and 40°C.

Download PDF (1.2 MB)

More emtest Manuals