Elcometer 138/2 Surface Contamination Kit User Guide

User guide for the Elcometer 138/2 Surface Contamination Kit, covering Bresle method sampling and pH, chloride and iron test strip measurements.


Elcometer 138/2 Surface Contamination Kit User Guide - cover page
Brand
Elcometer
Category
Other products
Document type
User Manual
Language
English
Pages
56
File format
PDF
File size
3 MB
Published
12 November, 2026
Updated
12 November, 2026
MD5 checksum
167246FF1CC21E7704E7D8A3EA92D00F

About this manual

The Elcometer 138/2 Surface Contamination Kit provides materials to collect a contamination sample from a surface in accordance with ISO 8502-6 (Bresle Method). A solution is produced from the test surface and then tested for pH, chloride concentration and iron concentration using single-use test strips.

The guide describes the two-stage test procedure, including producing the solution with a Bresle patch, testing with chloride, iron and pH strips, recording temperature, cleaning after test, technical specifications, spares and accessories, and legal notices.

What's inside

  • Overview
  • Box Contents
  • Test Procedure
  • Producing a Solution
  • Testing for Chloride
  • Testing for Iron
  • Testing for pH
  • Technical Specification
  • Spares & Accessories
  • Legal Notices & Regulatory Information

Specifications

Kit dimensions346 x 292 x 84 mm (13.6 x 11.5 x 3.3 in)
Kit weight2.1 kg (4 lb 7 oz)
Bresle patch size5 cm x 5 cm
Sample volume3 ml
Chloride measuring range30 to 600 ppm Cl (0.005-0.1% NaCl)
Iron measuring range3, 10, 25, 50, 100, 250, 500 mg/l Fe
pH measuring range0 to 14

Frequently asked questions

What does the Elcometer 138/2 kit measure?

It provides materials to collect a contamination sample and measure pH, chloride concentration and iron concentration using single-use test strips.

How much pure water is used to produce the solution?

Approximately 3 ml of pure water is injected into the Bresle patch.

How long should the solution remain in the patch?

For a suitable period of time; on un-pitted blast-cleaned areas, 10 minutes has been found satisfactory.

What should be done if the iron test strip color is too dark?

Repeat the measurement using fresh samples diluted with pure water until a value below 500 mg/l Fe is obtained, then multiply by the dilution factor.